ABSTRACT |
|---|
|
|
|---|
The application of left ventricular pressure-volume
analysis to transgenic mice to characterize the cardiac phenotype has been problematic due to the small size of the mouse heart and the rapid
heartbeat. Conductance technology has been miniaturized for the mouse
and can solve this problem. However, there has been no validation of
this technique. Accordingly, we performed echocardiography followed by
simultaneous ultrasonic crystals, flow probe, and conductance studies
in 18 CD-1 mice. Raw conductance volumes were corrected for an
inhomogenous electrical field (
) and parallel conductance
(Gpi) yielding a stroke volume of 14.1 ± 3.7 µl/beat, end-diastolic volume of 20.8 ± 6.5 µl, and
end-systolic volume of 9.0 ± 5.8 µl. The mean conductance
volumes were no different from those derived by flow probe and
echocardiography but did differ from ultrasonic crystals.
Gpi was determined to be 14.9 ± 8.7 µl.
However, hypertonic saline altered dimension and pressure in the mouse
left ventricle. Although Gpi can be determined
by the hypertonic saline method, saline altered hemodynamics,
questioning its validity in the mouse. Although mean measures of
absolute volume may be similar among different techniques, individual
values did not correlate.
conductance catheter; pressure-volume analysis; ultrasonic crystals; electrocardiography; transgenic mouse
| |
INTRODUCTION |
|---|
|
|
|---|
TO BE ABLE TO DEVELOP molecular targets for gene therapy in the treatment of patients with cardiac disease, we need to determine the phenotype that results from specific genetic manipulations. Transgenic mice provide a valuable way to relate specific gene products to phenotype; unfortunately, the ability to assess cardiovascular phenotype has presented a significant impediment (7, 10, 24). Although pressure-volume analysis to examine cardiac performance in the pressure-volume plane is possible in larger animals, larger transgenic animals are not available. The application of standard pressure-volume analysis methods to mice is not possible due to the small size of the mouse heart and rapid beat. Creating technology to generate an accurate, instantaneous volume signal in the transgenic mouse and beat-by-beat pressure-volume relationships in real time as cardiac dynamics change is critical for assessment of phenotype.
Measurement of electrical conductance by an indwelling catheter (conductance technology) offers a method to generate an instantaneous left ventricular (LV) volume signal in the mouse. The electric field generated from electrodes at the apex and immediately above the left ventricle is sensed by voltage measurement in the left ventricle as it fills and ejects blood. Although this technique has been utilized by previous investigators (17, 18), it has not been previously validated. We developed a signal generator/processor and bridge amplifiers that have been modified and licensed for commercial release (Mouse Conductance System-100, Millar Instruments, Houston, TX). When combined with a miniaturized mouse conductance catheter (SPR-719, Millar Instruments), a complete commercial system is now available to generate LV pressure-volume loops. Whereas this technology offers the ability to link physiology and molecular biology, no validation studies of this commercial technology have been published to date. Accordingly, the present study was performed to validate the currently commercially available conductance system in the mouse. The purpose of this study was to 1) establish normal values to serve as a reference for future studies in genetically altered mice; 2) examine the validity of current methods in determining parallel conductance; and 3) to compare this system to alternative technologies, including echocardiography and ultrasonic crystals.
| |
METHODS |
|---|
|
|
|---|
Animal Preparation
The protocol was approved by the Institutional Animal Care and Use Committee at the University of Texas Health Science Center in San Antonio and conformed with Guidelines for the Care and Use of Laboratory Animals (NIH Publication No. 86-23, Revised in 1985) and "Principles of Laboratory Animal Care" (published by the National Society for Medical Research). CD-1 mice (n = 18) weighing 20-30 g were anesthetized by administering methoxyflurane (Metofane, Pitman-Moore, Mundelein, IL) in a closed chamber, followed by the administration of urethan (1,000 mg/kg ip) and etomidate (25 mg/kg ip). Echocardiography was performed as outlined below before the chest was opened and the animal was placed on mechanical ventilation.Respiration was controlled through a tracheostomy cannula, and the animals were mechanically ventilated with room air at 60 breaths/min using a rodent ventilator (Harvard Apparatus model 680, South Natick, MA). The chest was entered via an anterior thoracotomy, and the descending thoracic aorta was dissected free above the level of the diaphragm. A small animal blood flowmeter (T 106, Transonic Systems, Ithaca, NY) was placed around the descending aorta. An apical stab was made with a 30-gauge needle, and the miniaturized mouse conductance catheter was advanced retrograde into the left ventricle along the cardiac long axis with the proximal electrode just within the myocardial wall of the LV apex. A pair of 1.0-mm piezoelectric crystals (Sonometrics, London, ON, Canada) were attached to the anterior and posterior walls of the left ventricle using cyanoacrylate adhesive (Vetbond, 3M Animal Care Products, St. Paul, MN). To determine parallel conductance with the hypertonic saline technique of Bann et al. (3), a 30-gauge needle was inserted into the right ventricle and left in place for the duration of the protocol. The needle was attached to a 4-mm length of polyethylene tubing-10 tubing connected to a 1-ml syringe containing 1.5% saline. At the termination of the experiment, the heart was removed from the chest cavity. The right ventricle, aorta, left atrium, mitral apparatus, and other structures were dissected away from the left ventricle. The heart was blotted, and the wet weight was determined (Electronic Balance).
Conductance System
A 1.4-Fr miniaturized pressure-conductance catheter (SPR-719, Millar Instruments, Houston, TX) was used in these studies. The catheter has four platinum electrodes, each 0.25 mm in length. Interelectrode spacing is 0.5, 4.5, and 0.5 mm between electrodes 1 and 2, 2 and 3, and 3 and 4, respectively. A constant excitation current (17 µA root mean square) and single frequency (10 kHz) were applied to the two outermost electrodes (electrodes 1 and 4) using a custom signal generator/processor and bridge amplifiers developed by us and subsequently modified (MCS-100, Millar Instruments). The results of the conductance studies in this paper also apply to the currently marketed ARIA Pressure Volume Conductance System (Millar Instruments) because the electronics for the first generation MCS-100 and the second generation ARIA are similar. The two inner electrodes (electrodes 2 and 3) were used to measure the instantaneous conductance signal. A high-fidelity pressure transducer was mounted between electrodes 2 and 3 to measure ventricular pressure (frequency response to 10 kHz).The theory behind the determination of volume using the conductance
catheter in larger animals has been described in detail elsewhere
(3, 5). Briefly, the tetrapolar catheter generates an
intraventricular electric field between the outer electrodes. The
potential differences between the inner electrode pair are measured
continuously and divided into current to yield instantaneous conductance. The time-varying ventricular volume
[Vi(t)] is calculated as
|
(1) |
is a volume calibration factor,
is the blood
resistivity, L is the distance between the sensing
electrodes (electrodes 2 and 3),
Gi(t) is the instantaneous
conductance, and Gpi is the instantaneous volume
of the surrounding structures, particularly the LV myocardium (i.e.,
parallel conductance).
System Calibration
The signal processor provides an inverted analog voltage output that when multiplied by the current is equal to the conductance detected by the measurement electrodes. A calibration procedure was developed to convert this conductance output (µS) into volume (µl) for each catheter and signal generator/processor. Six cylindrical holes ranging in size from 2- to 5-mm diameter and 15-mm deep hole were drilled in a Plexiglas block. The holes were filled with heparinized mouse blood at 25°C. The miniaturized mouse conductance catheter was suspended in the center of the hole, and the output conductance was digitally acquired and measured. Assuming a homogenous electrical field and no Gpi, Eq. 1 simplifies to
|
(2) |
|
(3) |
13.678, with
r2 = 0.9905, P < 0.0001)
was used to convert measured conductance to volume.
We also calculated Gpi to convert the
conductance signal to true volume according to Eq. 1. For
this purpose, the hypertonic saline technique described by Baan et al.
(3) was modified for the mouse. A 0.02-ml bolus of
hypertonic saline (1.5%) was injected into the right ventricle as
described above. As the saline entered the left ventricle following
transit through the lungs, it transiently increased blood conductivity.
This theoretically results in an increase in ventricular conductance
and an apparent increase in end-systolic (ESV) and end-diastolic volume
(EDV) without a true change in LV volume. ESV and EDV are related
during saline washin
|
(4) |
|
(5) |
Effect of Hypertonic Saline On Hemodynamics
We also tested the assumption that hypertonic saline bolus injection would not have an effect on LV hemodynamics (3). LV anterior to posterior short-axis dimension, LV end-systolic and end-diastolic pressures, and LV peak rate of pressure development (+dP/dt) and peak
dP/dt were evaluated via
digital ultrasonic crystals (Sonometrics) and micromanometer catheter
(SPR-719, Millar Instruments), respectively. Data from immediately
before the change in LV conductivity to the peak change in conductivity
as detected from the conductance catheter were compared.
Calculations
Conductance-derived pressure-volume (P/V) data were analyzed with customized software (PVAN). End-systolic pressure and volume were defined at the point of maximal P/V ratio, and end-diastolic pressure and volume were defined at the peak of the R wave on the electyrocardiogram. Stroke volume (SV) was defined as EDV
ESV,
and end-systolic elastance was determined from end-systolic pressure-volume relationships using the method of Suga et al. (35) to identify the point of maximal P/V ratio using an
iterative algorithm, where Pes = Ees(Ves
Vo),
where Pes is end-systolic pressure,
Ees is end-systolic elastance, Ves
is end-systolic volume, and Vo is the volume intercept.
Echocardiography
Equipment. A SONOS 5500 echocardiographic system (Hewlett-Packard, Andover, MA) was used to image the mouse hearts. Two-dimensional guided M mode imaging data were acquired at 1,000 lines/s using a 12-MHz phased array pediatric transducer operating at a frame rate of 70 Hz. The scrolling memory function built into the machine was used to select the best quality images for the M mode data. All data were recorded on commercial VHS video tape and printed with a Sony video printer (Sony, Japan) on 7.5 × 10 cm prints.
Imaging technique. Just before echocardiography, each animal was anesthetized as outlined above. The anterior chest wall was shaved, and the mouse was then secured to a custom-made water bed in a shallow left lateral decubitus position to facilitate ultrasound imaging. The animal was imaged in the parasternal short-axis plane through the anterior chest. The finger of a latex glove was filled with acoustic coupling gel (Aquasonic 100, Parker, Orange, NJ). The glove tip was centrifuged at 2,000 g to remove air bubbles contained within the gel and then placed over the tip of the transducer. This improved image quality by providing a 1.0- to 1.5-cm standoff between the transducer and the chest wall.
Electrocardiographic electrode clips were placed on the limbs for recording heart rate and timing intracardiac events using the single-channel electrocardiogram available on the echo machine. After two-dimensional imaging, a single M mode line was directed across the midventricular short-axis view perpendicular to the anterior and posterior walls. Position was verified using the two-dimensional image updated every few seconds. Analysis was performed on-line using the built-in analysis package.Data Analysis
All measurements were made using leading-edge conventions specified by the American Society of Echocardiography for clinical use in humans (41). Systole was defined as the smallest cavity diameter and diastole as the largest. LV volume was calculated using a prolate ellipse model assuming that the major axis was twice the minor axis. The rationale and limitations of this method have been reviewed and well validated in humans without regional wall motion abnormalities (9).Determination of Corrected SV
The conductance catheter creates an inhomogenous electrical field and therefore underestimates SV (44). The conductance signal has to be multiplied by a correction factor (ratio of true SV to conductance SV). To determine true SV, a small animal blood flowmeter (T 106, Transonics Systems) was placed around the descending thoracic aorta immediately above the level of the diaphragm. It could not be placed on the ascending aorta for technical reasons. To correct for the SV not passing through the ascending aorta, we determined the percentage of SV in the descending thoracic aorta as outlined below.A total of five CD-1 mice were used to determine true SV. This method has been previously described in part (12). A pediatric short focal length 12-MHz transducer (SONOS 5500, Hewlett-Packard) was used for cardiac imaging. SV was determined in both the ascending (sinotubular junction) and descending aorta (at the left subclavian artery level). SV was determined noninvasively from pulsed wave aortic Doppler recordings (gate length 0.06 cm without angle correction) and two-dimensional measurements of the diameter of the proximal ascending and descending aorta. All echocardiographic data were obtained from the suprasternal approach.
SV was determined by multiplying the flow velocity integral of the
pulsed-Doppler recording of each beat by the respective aortic
cross-sectional area (CSA). CSA was calculated assuming a circular
orifice with the following formula: CSA =
× (Aod)2/2, where Aod is the aortic
outer diameter. Cardiac output was then calculated as the
product of SV and heart rate.
Digital Ultrasonic (Crystal) Measurement System
A pair of 1.0-mm piezoelectric crystals (Sonometrics) attached to the anterior and posterior walls of the left ventricle provided an instantaneous, short-axis, epicardial diameter. Intraventricular volume was derived from this signal by first converting it to an outer volume of the LV (LVo) by means of the standard formula for a prolate ellipse, assuming that the major axis is twice the minor axis
|
(6) |
Statistics
Separate analyses were performed for SV, EDV, and ESV using repeated measures analysis of variance (ANOVA), followed by pair-wise Newman-Keuls multiple comparisons. Pearson correlation coefficients were used to compare LV weight and Gpi determined with the hypertonic saline method. All computations were performed using SAS (version 6.11, SAS Institute, Cary, NC).| |
RESULTS |
|---|
|
|
|---|
In five CD-1 mice examined by echocardiography and descending
aortic flow probe, we assessed the fraction of SV distributed to the
descending aorta. In these mice, 72.5 ± 4.8% of the SV ejected
into the ascending thoracic aorta was detected in the desending
thoracic aorta. In subsequent studies, we used this value to correct
for total SV when flow probe data were used to calibrate conductance
data. Thus the SV-corrected data shown in Table
1 are adjusted to reflect total cardiac
output.
|
Table 1 is a summary of hemodynamic and conductance parameters for 13 mice studied. The baseline heart rate was 426 ± 52 beats/min. The
LV end-systolic pressure for our open-chest mice was 57 ± 16 mmHg. The descending thoracic aorta flow was 10.1 ± 2.3 µl/beat. Proper determination of
requires ascending aorta flow.
Therefore, the flow probe data were normalized by 72.5% as described
above. The corrected SV determined from the flow probe was 13.9 ± 3.1 µl/beat.
In this group of mice,
was 1.4 ± 0.5, and
Gpi was 14.9 ± 8.7 µl. The
Gpi was large in comparison to the raw
conductance EDV of 28.0 ± 8.2 µl and ESV of 19.3 ± 8.1 µl. When the raw conductance signals were corrected for
and
Gpi, the EDV was 20.8 ± 6.5 µl, the ESV
was 9.0 ± 5.8 µl, and the SV was 14.1 ± 3.7 µl/beat.
Increase in Gpi With LV Mass
Figure 1 is a plot of calculated Gpi as a function of LV mass (in mg) determined directly after the left ventricle was excised and weighed. There was a correlation between Gpi determined by the hypertonic saline method and LV mass (r = 0.74, P = 0.006).
|
Caval Occlusion
There were appropriate decreases in volume and dimension with occlusion of the inferior vena cava for both the conductance catheter and epicardial piezoelectric crystals. Figure 2 is example data from a single representative mouse. There are beat-by-beat decreases in simultaneous conductance volume and crystal dimension. The conductance-derived end-systolic elastance for this mouse was 10.62 mmHg/µl and Vo was 6.22 µl. The crystal data are discontinuous because the magnitude of wall motion in these small hearts is near the absolute level of resolution for this instrument.
|
SV
Eight of thirteen mice had complete crystal data. Of the five mice that were excluded, two had negative ESV and EDV calculated from Eq. 6 in the METHODS, and three had technical errors. These mice were excluded based on inadequate crystal data. The conductance and echo data were usable. Therefore, the following ANOVA were performed on the eight mice with data common to all techniques.SV values determined by corrected flow probe (means ± SD = 13.9 ± 3.1 µl/beat), conductance (14.1 ± 3.7 µl/beat),
and echocardiography (14.1 ± 4.9 µl/beat) were not
significantly different by repeated measures ANOVA (P = 0.84). A Newman-Keuls multiple comparisons procedure indicated that
crystal SV (10.6 ± 1.7 µl/beat) was significantly less than any
of the other three measurements (P < 0.05). Figure 3 is a plot of SV using the conductance
technique, echocardiography, and piezoelectric crystals versus SV
defined by the corrected flow probe. Despite similar mean conductance,
echocardiography, and flow probe-derived SV, for a given mouse there
are dissimilarities between the methods examined. There is an apparent
correlation for individual mouse SV between the flow probe and the
conductance catheter technique, but this is due to the use of the flow
probe to derive
and correct the raw conductance signal.
|
EDV
EDV values determined by corrected conductance (20.8 ± 6.5 µl) and echocardiography (18.2 ± 7.5 µl) were not significantly different (P = 0.41), but EDV defined from ultrasonic dimension gauges (38.7 ± 8.9 µl) were elevated compared with the other techniques (P < 0.01). Figure 4 is a plot of EDV using echocardiography and piezoelectric crystals versus EDV defined by the conductance catheter. Despite similar mean conductance and echocardiographic EDV, for a given mouse there is variability between the methods examined.
|
ESV
ESV values determined by corrected conductance (9.0 ± 5.8 µl) and echocardiography (4.1 ± 5.6 µl) were not significantly different (P = 0.17), but piezoelectric crystal ESV (32.1 ± 13.1 µl) was significantly elevated compared with both other procedures (P < 0.01). Figure 5 shows a plot of ESV using echocardiography and piezoelectric crystals versus ESV defined by the conductance catheter. Despite similar mean conductance and echocardiographic ESV, for a given mouse there is variability between the methods examined.
|
Effect of Hypertonic Saline on Hemodynamics
The use of the hypertonic saline method to determine Gpi is based on the assumption that LV blood conductivity is changing but not LV hemodynamics (8). Table 2 examines whether this assumption is true with saline washin data from LV epicardial piezoelectric crystal dimensions and LV pressures. There is a significant increase in end-diastolic and end-systolic dimension from immediately before the LV saline washin phase (baseline) to the peak of data acqusition for the determination of Gpi (saline). There is also a significant increase in LV end-systolic and end-diastolic pressures (Fig. 6) and in +dP/dt and
dP/dt over
this same time.
|
|
Figure 7 is an example of simultaneous
data from an individual mouse. Figure 7A demonstrates the
calculation of Gpi using Eqs. 4 and 5 in the METHODS. Gpi was
12.48 µl. Figure 7B is the conductance-derived LV
pressure-volume relationship during hypertonic saline washin used to
calculate ESV and EDV on a beat-by-beat basis shown in Fig.
7A. Figure 7C is the LV pressure-dimension relationship at the same time demonstrating that dimension and pressure
were increasing on a beat-by-beat basis.
|
| |
DISCUSSION |
|---|
|
|
|---|
The present study provides reference values in mice for the
miniaturized mouse conductance system. This includes reference values
for leakage of the electrical field into the myocardium (Gpi) and correction of the inhomogenity of the
electrical field in the left ventricle (
). Using these reference
values, we have determined for the first time the absolute ESV, EDV,
and SV using conductance technology in the mouse. We then examined the
validity of determining Gpi with the standard
hypertonic saline techniques used in larger animals (4, 27,
36-38, 40, 42), and we found that a steady-state
hemodynamic condition was not maintained while the hypertonic saline
traveled through the left ventricle. Given the assumptions of this
approach known to be valid in larger animals (3), this
provides a substantial limitation of this approach for its use in mice.
Finally, we compared the absolute LV volumes obtained with other
independent technologies that also have the capability to generate LV
pressure-volume or dimension relationships. Although similar mean
results were obtained among echocardiography, the conductance
technique, and the flow probe, there were poor correlations for
individual mice for EDV and ESV for the conductance technique,
echocardiography (r2 = 0.119, r2 = 0.094, respectively) and ultrasonic
crystals (r2 = 0.028, r2 = 0.066, respectively).
The three techniques examined, the conductance technique, echocardiography, and ultrasonic crystals, were chosen because all have the capability to generate instantaneous LV pressure-volume or dimension relationships (8, 17, 18, 26, 30). Review of current published studies with mice report the use of numerous echocardiographic measures of LV function that are known to be load dependent (15, 20, 22, 31, 34, 39). More load-independent reports of stress-shortening or stress-velocity relationships (12, 21, 43) using echocardiography have been successfully applied to the in vivo mouse left ventricle. However, two-dimensional echocardiography suffers from transducers of which the typical imaging rate is 30 Hz, which is too slow for mouse hearts whose rate exceeds 5 Hz, making these studies dependent on M mode at 1,000 Hz. This prevents imaging the mouse LV long axis and the subsequent substitution of meridional for circumferential wall stress in these studies. Finally, the use of pharmacological agents to alter steady-state load (such as phenylephrine or methoxamine) may have direct myocardial effects (13), which alters the value of this technique as a way to define contractility. The use of M mode echocardiography also precludes the application of stress-shortening relationships to ventricles with wall motion abnormalities, which is not a limitation of the conductance technique.
Other investigators have applied ultrasonic epicardial crystals to generate LV pressure-dimension relationships (26, 30). This approach is limited by the size of the heart, because the LV wall thickness of the mouse is roughly the same thickness as the crystals. The result is an offset of the crystals above the epicardium. Because EDV and ESV are derived from the cube of crystal dimension minus LV mass, EDV and ESV will always be overestimated compared with the conductance and echocardiographic results as we found. In addition, the majority of LV thickening occurs on the endocardial surface (14, 33) and not the epicardial surface where the crystals are located. Therefore, epicardial crystals would be anticipated to underestimate SV compared with endocardial methods such as conductance or echocardiography, as we found. Finally, because ultrasonic crystals work by transmitting and receiving short pulses of ultrasonic energy, there is a finite limit of temporal resolution inherent in the technique. This resolution limit is most evident when absolute motion is small as is visually evident in Figs. 2 and 7, where simultaneous conductance and ultrasonic crystal data are shown. The discontinuous appearance of the crystal signal results from this phenomena and may be particularly relevant in the mouse due to its rapid heart rate. Although the use of smoothing algorithms will improve this, there will be a small loss of precision in the measurement. A similar occurrence is not evident in the conductance signal, because the voltage change occurs instantaneously, and analog-to-digital conversion occurs at very high rates. Some of these limitations of ultrasonic crystals may be overcome with endocardial placement, an approach that has recently been used successfully (11).
Conductance technology seems ideally suited to generate pressure-volume relationships in the mouse left ventricle. The electrical field generated by the catheter is independent of LV chamber geometry, and therefore mice with hypertrophy or regional wall motion abnormalities can be accurately assessed. The conductance catheter is measuring an analog voltage output, which varies instantaneously with the volume of the left ventricle, and therefore provides a continuous LV volume signal that is accurate despite the rapid heart rate of the mouse. The accuracy of conductance technology in determining absolute volumes and SV has been established previously in in vivo rat (23), canine (1, 2, 5), and human hearts (3). In the current study, we extend these findings to the mouse and have shown that the mean SV for a group of eight mice, determined by the conductance method, is comparable to the mean SV defined by flow probe or echocardiography. Likewise, we found that the mean conductance EDV and ESV are comparable to echocardiography.
Whereas we found similar mean values for conductance, flow probe, and echocardiographic SVs and similar mean values for conductance and echocardiographic ESV and EDV for a given mouse, there were discrepancies between the methods examined. It is important to note that each of these methods has the capability to detect similar directional changes in instantaneous volume or dimension. Figure 2 shows simultaneous data from epicardial crystals and an intraventricular conductance catheter: as the inferior cava is occluded, both crystal dimension and conductance volume fall on a beat-by-beat basis. Therefore, each technique is internally consistent for the detection of acute hemodynamic changes. However, individual animals did not correlate, likely due to limitations inherent in each technique, which will vary from mouse to mouse. For instance, echocardiography has a variable ability to define the endocardial border. Crystals have a variable offset due to their placement on the epicardium. Conductance has a variable ability to determine parallel conductance. Because these variances operate randomly between animals, it is not surprising that specific indexes do not correlate well, despite concurrent average results. Importantly, however, each technique is consistent. The choice of which of these three methods to utilize will therefore vary with the particular study design. Echocardiography is ideal for measurement of LV thickness, morphology, and mass; crystals for segmental function; and conductance for determination of an instantaneous volume change during inferior vena cava occlusion.
The electrical field generated by the conductance catheter leaks into the LV myocardium. Therefore, the instantaneous volume signal is a combination of LV blood volume and myocardium. The portion of this signal not arising from the LV blood volume has been termed Gpi. The use of bolus injection of hypertonic saline to define Gpi has been validated by previous investigators in larger animals (4, 27, 36-38, 40, 42). Of critical importance to the derivation of Gpi by this method is the assumption that the saline has no significant effect on LV volume or performance (3). In larger animals the typical volume injected is 1 ml in the canine or 10 ml in the human and represents <2% of the LV blood volume. In the mouse, the injection of 20 µl of hypertonic saline represents nearly the entire LV blood volume (Table 1), an amount that we found to cause significant hemodynamic changes in LV pressure, dP/dt, and dimension (Table 2). Previous investigators (6) have similarly shown errors in small mammals with the saline technique. Alternative techniques to quantify Gpi such as the use of more than one input frequency (16) need to be developed. Despite these limitations of the accuracy of the hypertonic saline technique, we were able to detect increased Gpi in mice with greater LV weight (Fig. 1) as would be anticipated. This has not been previously demonstrated in larger animals.
Another limitation of this study is that an open-chest preparation was used. This approach lowers blood pressure and heart rate (25) and is terminal. Such an approach raises issue with using conductance technology to study more physiological hemodynamics, as found in the conscious mouse, and with making serial studies in precious transgenic mice. One solution is the use of the carotid artery to cannulate the left ventricle. The inferior vena cava can be accessed from a subxyphoid incision. This approach offers the possibility that the mouse could survive an initial study. In addition, the use of 100% oxygen via mouse ventilators and cannulation of the internal jugular vein for fluid, colloid, or blood administration were not utilized in the present study and may have improved hemodynamics.
A final limitation is that hypertonic saline overestimated Gpi by altering hemodynamics. The injection of 20 µl of saline into the right ventricle represents a large portion of LV blood volume. This leads to significant beat-to-beat changes in EDV greater than ESV as demonstrated in Fig. 7B. This is expected because the bolus of hypertonic saline produces a sudden increase in preload without a change in contractility. To determine Gpi, Ves is plotted on the ordinate and Ved is plotted on the abscissa as shown in Fig. 7A, and Vp is determined as the intercept of the extrapolated data with the line of identity. Because Ved is changing faster than Ves, violating the assumptions of Baan et al. (3), the intersection with the line of identity will be artificially increased. The result will be overestimation of Vp, and smaller calculated EDV and ESV determined with the conductance system.
In conclusion, the miniaturized mouse conductance system can generate an instantaneous volume signal for the study of cardiovascular performance in the pressure-volume plane in very small hearts. This is important because genetic alterations will allow definitive structure-function studies of the cardiovascular system. Because selected mutations may affect not only the molecular structure of cardiac myocytes, but also the vasculature, altering both contractility and loading conditions, the ability to evaluate alterations in contractility in vivo independent of simultaneous changes in loading conditions will allow a better physiological understanding of the cardiovascular phenotype of gene-altered mice.
| |
ACKNOWLEDGEMENTS |
|---|
We thank Danny Escobedo for excellent surgical assistance. Dr. Gary M. Clark, in the Biostatistics and Data Processing Unit, performed statistics at the University of Texas Health Science Center in San Antonio, with support from the Cancer Center Support Grant (PO-CA54174). Karen Bussolati of Hewlett-Packard loaned the SONOS 5500 Echocardiography equipment for the echocardiography performed in San Antonio, TX.
| |
FOOTNOTES |
|---|
This study was supported by the University of Texas Health Sciences Center in San Antonio (to M. D. Feldman), the Department of Veterans Affairs (to M. D. Feldman and G. L. Freeman), and a research grant from Millar Instruments, Houston, TX (to M. D. Feldman).
This study was presented in part at the 1999 American Heart Association Scientific Sessions in Atlanta, Georgia.
Address for reprint requests and other correspondence: M. D. Feldman, Rm. 5.642U, Univ. of Texas Health Science Center in San Antonio, 7703 Floyd Curl Dr., San Antonio, TX 78284-7872 (E-mail: feldmanm{at}uthscsa.edu).
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
Received 14 December 1999; accepted in final form 10 April 2000.
| |
REFERENCES |
|---|
|
|
|---|
1.
Applegate, RJ,
Cheng CP,
and
Little WC.
Simultaneous conductance catheter and dimension assessment of left ventricle volume in the intact animal.
Circulation
81:
638-648,
1990
2.
Baan, J,
Jong TT,
Kerkhof PL,
Moene RJ,
Van Dijk ET,
Van der Velde ET,
and
Koops J.
Continuous stroke volume and cardiac output from intraventricular dimensions obtained with impedance catheter.
Cardiovasc Res
15:
328-334,
1981[ISI][Medline].
3.
Baan, J,
Van der Velde ET,
De Bruin HG,
Smeenk GJ,
Koops J,
Van Dijk AD,
Temmerman D,
Senden J,
and
Buis B.
Continuous measurement of left ventricular volume in animals and humans by conductance catheter.
Circulation
70:
812-823,
1984
4.
Boltwood, CM,
Appleyard RF,
and
Glantz SA.
Left ventricular volume measurement by conductance catheter in intact dogs: parallel conductance volume depends on left ventricular size.
Circulation
80:
1360-1377,
1989
5.
Burkhoff, D,
Van der Velde E,
Kass D,
Baan J,
Maughan WL,
and
Sagawa K.
Accuracy of volume measurement by conductance catheter in isolated, ejecting canine hearts.
Circulation
72:
440-447,
1985
6.
Cassidy, SC,
and
Teitel DF.
The conductance volume catheter technique for measurement of left ventricular volume in young piglets.
Pediatr Res
31:
85-90,
1992[ISI][Medline].
7.
Christensen, G,
Wang Y,
and
Chien KR.
Physiological assessment of complex cardiac phenotypes in genetically engineered mice.
Am J Physiol Heart Circ Physiol
272:
H2513-H2524,
1997
8.
Denault, AY,
Gorcsan J,
Mandarino WA,
Kancel MJ,
and
Pinsky MR.
Left ventricular performance assessed by echocardiographic automated border detection and arterial pressure.
Am J Physiol Heart Circ Physiol
272:
H138-H147,
1997
9.
Deveroux, RB,
and
Reichek N.
Echocardiographic determination of left ventricular mass in man: anatomic validation of the method.
Circulation
55:
613-618,
1997
10.
Doevendans, PA,
Daemen MJ,
de Muinck ED,
and
Smits JF.
Cardiovascular phenotyping in mice.
Cardiovasc Res
39:
34-49,
1998
11.
Esposito, G,
Defer N,
Mao L,
Hanoune J,
and
Rockman HA.
Myocardial contractile function in transgenic mice overexpressing adenylyl cyclase 8 as assessed by the end-systolic pressure volume relation (Abstract).
Circulation
100:
I-760,
1999.
11a.
Feldman, MD,
Mao Y,
Valvano JW,
Pearce JA,
and
Freeman GL.
Development of a multifrequency conductance catheter-based system to determine LV function in mice.
Am J Physiol
279:
H1411-H1420,
2000
12.
Frentzke, RC,
Korcarz CE,
Shroff SG,
Lin H,
Sandelski J,
Leiden JM,
and
Lang RM.
Evaluation of ventricular and arterial hemodynamics in anesthetized closed-chest mice.
J Am Soc Echocardiogr
10:
915-925,
1997[ISI][Medline].
13.
Freeman, GL,
Little WC,
and
O'Rourke RA.
The effect of vasoactive agents on the left ventricular end-systolic pressure-volume relation in closed chest dogs.
Circulation
74:
1107-1113,
1986
14.
Gallagher, KP,
Osakada G,
Matsuzaki M,
Miller M,
Kemper WC,
and
Ross J, Jr.
Nonuniformity of inner and outer systolic wall thickening in conscious dogs.
Am J Physiol Heart Circ Physiol
249:
H241-H248,
1985.
15.
Gardin, JM,
Siri FM,
Kitsis RN,
Edwards JG,
and
Leinwand LA.
Echocardiographic assessment of left ventricular mass and systolic function in mice.
Circ Res
76:
907-914,
1995
16.
Gawne, TJ,
Gray KS,
and
Goldstein RE.
Estimating left ventricular offset volume using dual-frequency conductance catheters.
J Appl Physiol
63:
872-876,
1987
17.
Georgakopoulos, D,
Christe ME,
Giewat M,
Seidman CM,
Seidman JG,
and
Kass DA.
The pathogenesis of familial hypertrophic cardiomyopathy: early and evolving effects from an
-cardiac myosin heavy chain missense mutation.
Nat Med
5:
327-330,
1999[ISI][Medline].
18.
Georgakopoulos, D,
Mitzner WA,
Chen CH,
Byrne BJ,
Millar HD,
Hare JM,
and
Kass DA.
In vivo murine left ventricular pressure-volume relations by miniaturized conductance micromanometry.
Am J Physiol Heart Circ Physiol
274:
H1416-H1422,
1998
19.
Haas, GJ,
McCune SA,
Brown DM,
and
Cody RJ.
Echocardiographic characterization of left ventricular adaptation in a genetically determined heart failure rat model.
Am Heart J
130:
806-811,
1995[ISI][Medline].
20.
Harada, K,
Komuro I,
Shiojima I,
Hayashi D,
Kudoh S,
Mizuno T,
Kazuhisa K,
Matsubara H,
Sugaya T,
Murakami K,
and
Yazaki Y.
Pressure overload induces cardiac hypertrophy in angiotensin II type 1A receptor knockout mice.
Circulation
97:
1952-1959,
1998
21.
Hoit, BD,
Khan ZU,
Pawloski-Dahm CM,
and
Walsh RA.
In vivo determination of left ventricular wall stress-shortening relationship in normal mice.
Am J Physiol Heart Circ Physiol
272:
H1047-H1052,
1997
22.
Hoit, BD,
Khoury SF,
Kranias EG,
Ball N,
and
Walsh RA.
In vivo echocardiographic detection of enhanced left ventricular function in gene-targeted mice with phospholamban deficiency.
Circ Res
77:
632-637,
1995
23.
Ito, H,
Takaki M,
Yamaguchi H,
Tachibana H,
and
Suga H.
Left ventricular volumetric conductance catheter for rats.
Am J Physiol Heart Circ Physiol
270:
H1509-H1514,
1996
24.
James, JF,
Hewett TE,
and
Robbins J.
Cardiac physiology in transgenic mice.
Circ Res
82:
407-415,
1998
25.
Kass, DA,
Hare JM,
and
Georgakopoulos D.
Murine cardiac function: a cautionary tail.
Circ Res
82:
519-522,
1998
26.
Kubota, T,
Mahler CM,
McTiernan CF,
Wu CC,
Feldman MD,
and
Feldman AM.
End-systolic pressure-dimension relationship of in situ mouse left ventricle.
J Mol Cell Cardiol
30:
357-363,
1998[ISI][Medline].
27.
Lankford, EB,
Kass DA,
Maughan WL,
and
Shoukas AA.
Does volume catheter parallel conductance vary during a cardiac cycle?
Am J Physiol Heart Circ Physiol
258:
H1933-H1942,
1990
28.
Litwin, SE,
Katz SE,
Morgan JP,
and
Douglas PS.
Serial echocardiographic assessment of left ventricular geometry and function after large myocardial infarction in the rat.
Circulation
89:
345-354,
1994
29.
Litwin, SE,
Katz SE,
Weinberg EO,
Lorell BH,
Aurigemma GP,
and
Douglas PS.
Serial echocardiographic-doppler assessment of left ventricular geometry and function in rats with pressure-overload hypertrophy.
Circulation
91:
2642-2654,
1995
30.
Miyamoto, MI,
Kim CS,
Guerrero JL,
Rosenzweig A,
Gwathmey JK,
and
Hajjar RJ.
Ventricular pressure and dimension measurements in mice.
Lab Anim Sci
49:
305-307,
1999[ISI][Medline].
31.
Patten, RD,
Aronovitz MJ,
Deras-Mejia L,
Pandian NG,
Hanak GG,
Smith JJ,
Mendelsohn ME,
and
Konstam MA.
Ventricular remodeling in a mouse model of myocardial infarction.
Am J Physiol Heart Circ Physiol
274:
H1812-H1820,
1998
32.
Pawlush, DG,
Moore RL,
Musch TI,
and
Davidson Jr WR.
Echocardiographic evaluation of size, function, and mass of normal and hypertrophied rat ventricles.
J Appl Physiol
74:
2598-2605,
1993
33.
Rademakers, FE,
Rogers WJ,
Guier WH,
Hutchins GM,
Siu CO,
Weisfeldt ML,
Weiss JL,
and
Shapiro EP.
Relation of regional cross-fiber shortening to wall thickening in the intact heart.
Circulation
89:
1174-1182,
1994
34.
Sakata, Y,
Hoit BD,
Liggett SB,
Walsh RA,
and
Dorn II GW.
Decompensation of pressure-overload hypertrophy in G
q-overexpressing mice.
Circulation
97:
1488-1495,
1998
35.
Suga, H,
and
Sagawa K.
Instantaneous pressure-volume relationships and their ratio in the excised, supported canine left ventricle.
Circ Res
35:
117-126,
1974
36.
Szwarc, RS,
and
Ball HA.
Simultaneous LV and RV volumes by conductance catheter: effects of lung insufflation on parallel conductance.
Am J Physiol Heart Circ Physiol
275:
H653-H661,
1998
37.
Szwarc, RS,
Laurent D,
Allegrini PR,
and
Ball HA.
Conductance catheter measurement of left ventricular volume: evidence for nonlinearity within cardiac cycle.
Am J Physiol Heart Circ Physiol
268:
H1490-H1498,
1995
38.
Szwarc, RS,
Mickleborough LL,
Mizuno SI,
Wilson GJ,
Liu P,
and
Mohamed S.
Conductance catheter measurements of left ventricular volume in the intact dog: parallel conductance is independent of left ventricular size.
Cardiovasc Res
28:
252-258,
1994
39.
Tanaka, N,
Dalton N,
Mao L,
Rockman HA,
Peterson KL,
Gottshall KR,
Hunter JJ,
Chien KR,
and
Ross J, Jr.
Transthoracic echocardiography in models of cardiac disease in the mouse.
Circulation
94:
1109-1117,
1996
40.
Teitel, DF,
Klautz RJM,
Cassidy SC,
Steendijk P,
van der Velde ET,
van Bel F,
and
Baan J.
The end-systolic pressure-volume relationship in young animals using the conductance technique.
Eur Heart J
13, SupplE:
40-46,
1992.
41.
Weyman AE. Cross-Sectional Echocardiography.
Philadelphia, PA: 1982, p. 289-297.
42.
White, PA,
Chaturvedi RR,
Shore D,
Lincoln C,
Szwarc RS,
Bishop AJ,
Oldershaw PJ,
and
Redington AN.
Left ventricular parallel conductance during cardiac cycle in children with congenital heart disease.
Am J Physiol Heart Circ Physiol
273:
H295-H302,
1997
43.
Williams, RV,
Lorenz JN,
Witt SA,
Hellard DT,
Khoury PR,
and
Kimball TR.
End-systolic stress-velocity and pressure-dimension relationships by transthoracic echocardiography in mice.
Am J Physiol Heart Circ Physiol
274:
H1828-H1835,
1998
44.
Wu, CC,
Skalak TC,
Schwenk TR,
Mahler CM,
Anne A,
Finnerty PW,
Haber HL,
Weikle RM, II,
and
Feldman MD.
Accuracy of the conductance catheter for measurement of ventricular volumes seen clinically: effects of electric field homogeneity and parallel conductance.
IEEE Trans Biomed Eng
4:
266-277,
1997.
This article has been cited by other articles:
![]() |
R. Bauer, G.A. MacGowan, A. Blain, K. Bushby, and V. Straub Steroid treatment causes deterioration of myocardial function in the {delta}-sarcoglycan-deficient mouse model for dilated cardiomyopathy Cardiovasc Res, September 1, 2008; 79(4): 652 - 661. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. M. Shioura, D. L. Geenen, and P. H. Goldspink Assessment of cardiac function with the pressure-volume conductance system following myocardial infarction in mice Am J Physiol Heart Circ Physiol, November 1, 2007; 293(5): H2870 - H2877. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Nielsen, S. B. Kristiansen, S. Ringgaard, T. T. Nielsen, A. Flyvbjerg, A. N. Redington, and H. E. Botker Left ventricular volume measurement in mice by conductance catheter: evaluation and optimization of calibration Am J Physiol Heart Circ Physiol, July 1, 2007; 293(1): H534 - H540. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Reyes, M. E. Steinhelper, J. A. Alvarez, D. Escobedo, J. Pearce, J. W. Valvano, B. H. Pollock, C.-L. Wei, A. Kottam, D. Altman, et al. Impact of physiological variables and genetic background on myocardial frequency-resistivity relations in the intact beating murine heart Am J Physiol Heart Circ Physiol, October 1, 2006; 291(4): H1659 - H1669. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Jegger, X. Jeanrenaud, M. Nasratullah, P.-G. Chassot, A. Mallik, H. Tevaearai, L. K. von Segesser, P. Segers, and N. Stergiopulos Noninvasive Doppler-derived myocardial performance index in rats with myocardial infarction: validation and correlation by conductance catheter Am J Physiol Heart Circ Physiol, April 1, 2006; 290(4): H1540 - H1548. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. E. Claessens, D. Georgakopoulos, M. Afanasyeva, S. J. Vermeersch, H. D. Millar, N. Stergiopulos, N. Westerhof, P. R. Verdonck, and P. Segers Nonlinear isochrones in murine left ventricular pressure-volume loops: how well does the time-varying elastance concept hold? Am J Physiol Heart Circ Physiol, April 1, 2006; 290(4): H1474 - H1483. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. L. Lindsey, G. P. Escobar, L. W. Dobrucki, D. K. Goshorn, S. Bouges, J. T. Mingoia, D. M. McClister Jr., H. Su, J. Gannon, C. MacGillivray, et al. Matrix metalloproteinase-9 gene deletion facilitates angiogenesis after myocardial infarction Am J Physiol Heart Circ Physiol, January 1, 2006; 290(1): H232 - H239. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. A. MacGowan, J. Rager, S. G. Shroff, and M. A. Mathier In vivo {alpha}-adrenergic responses and troponin I phosphorylation: anesthesia interactions J Appl Physiol, April 1, 2005; 98(4): 1163 - 1170. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Pacher, J. G. Mabley, L. Liaudet, O. V. Evgenov, A. Marton, G. Hasko, M. Kollai, and C. Szabo Left ventricular pressure-volume relationship in a rat model of advanced aging-associated heart failure Am J Physiol Heart Circ Physiol, November 1, 2004; 287(5): H2132 - H2137. [Abstract] |