1 Hydraulics Laboratory, Institute of Biomedical Technology, University of Gent, 9000 Gent, Belgium; 2 The Cleveland Clinic Foundation, Cleveland, Ohio 44195; 3 Norwegian University of Science and Technology, Trondheim, Norway; and 4 Cardiology Department, University Hospital, University of Gent, Belgium
| |
ABSTRACT |
|---|
|
|
|---|
We fitted a
three-segment transmission line model for the radial-carotid/aorta
pressure transfer function (TFF) in 31 controls and 30 patients with
coronary artery disease using noninvasively measured (tonometry) radial
and carotid artery pressures (Pcar). Except for the distal
reflection coefficient (0.85 ± 0.21 in patients vs. 0.71 ± 0.25 in controls; P < 0.05), model parameters were not
different between patients or controls. Parameters were not related to
blood pressure, age, or heart rate. We further assessed a
point-to-point averaged TFF (TFFavg) as well as upper
(TFFmax) and lower (TFFmin) enveloping TFF.
Pulse pressure (PP) and augmentation index (AIx) were derived on
original and reconstructed Pcar (Pcar,r). TFFavg yielded closest morphological agreement between
Pcar and Pcar,r (root mean square = 4.3 ± 2.3 mmHg), and TTFavg best predicted PP
(41.5 ± 11.8 vs. 41.1 ± 10.0 mmHg measured) and AIx
(
0.02 ± 0.19 vs. 0.01 ± 0.19). PP and AIx, calculated
from Pcar or Pcar,r, were higher in patients
than in controls, irrespectively of the TFF used. We conclude that
1) averaged TFF yield significant discrepancies between
reconstructed and measured pressure waveforms and subsequent derived
AIx; and 2) different TFFs seem to preserve the information in the pressure wave that discriminates between controls and patients.
blood pressure; model; applanation tonometry
| |
INTRODUCTION |
|---|
|
|
|---|
KNOWLEDGE OF CENTRAL AORTA PRESSURE is clinically important because it permits, together with aortic flow, one to compute arterial input impedance and hemodynamic parameters characterizing ventricular-arterial interaction (wave reflection indexes, hydraulic power) of diagnostic value in cardiology (1, 5, 14, 19). Using tonometry, one can estimate, noninvasively, the central aortic pressure wave by measuring the arterial pulse 1) at a superficial artery close enough to the heart (subclavian, carotid artery) so that the effects of wave distortion can be neglected (1, 5, 6, 14, 16), or 2) at a peripheral superficial artery (radial or carotid artery) with use of a pressure transfer function (TFF) to compute the pressure wave at the central aorta (4, 7-9, 12).
Averaging transfer functions, measured directly in (relatively) large human populations, yielded generalized TFF for the radial-aorta (7, 12) and the carotid-aorta (11) pathway. Such generalized TFF have been integrated into commercially available systems, predicting central aorta pressure and derived indexes, such as the augmentation index (AIx), from the carotid or radial artery pressure wave (4, 23). However, there is large scatter in measured transfer functions, and the morphology of reconstructed central pressure, using such a generalized TFF, may differ considerably from the directly measured pressure.
The aim of this work was to provide a method to individualize TFF, based on a transmission line model for the radial-aortic/carotid pathway (9). If model parameters characterizing the pressure transfer function are related to easily measured patient characteristics (age, blood pressure, etc.), then TFF can be predicted on a patient-to-patient basis, enabling a better estimate of the central aorta pressure wave morphology. First, we fitted the model to a published radial-aorta pressure transfer function that was derived from invasively recorded pressures (7). This way, we obtained reference values for all model parameters. Second, we fitted the model to (noninvasively) measured radial-carotid transfer functions in controls and in patients using a selected number of model parameters, and we studied the correlation between these model parameters and easily measured hemodynamic indexes. Because we were unable to assess any relation, we further studied the impact of using upper and lower limit transfer functions, enveloping measured data, on estimated central blood pressure and on derived indexes such as pulse pressure (PP) and the AIx.
| |
MATERIALS AND METHODS |
|---|
|
|
|---|
Transmission Line Model For Radial-Aortic/Carotid Pathway
The model simulating the aorta-radial pathway is analoguous to the model used by Karamanoglu and Fenely (9) for the finger carotid pathway and consists of a stepwise tapered transmission line with three segments (proximal, middle, and distal) terminated by a lumped parameter model (Fig. 1). For each segment, the transmission line equations describing the propagation of pressure (P) and flow (Q) harmonics waves apply
|
), accounting for
wave propagation and damping, is given as
|
|
as a function of frequency (
) (2,
9, 10)
|
|
= 1,050 kg/m3. Values for vessel
dimensions and
0 are given in Table
1. The impedance mismatch between the
terminating windkessel model and the distal segment results in wave
reflection. Instead of explicitly modeling the terminating lumped
parameter model, the distal reflection coefficient (
= Pb/Pf =
Qb/Qf) is modeled as
(9)
|
0 and
are the modulus and phase of the
distal reflection coefficient, respectively.
|
|
Assessing Reference Model Parameters: Fitting the Model to the Transfer Function of Chen et al.
The modulus and phase angle data for the steady-state radial-aorta transfer function of Chen et al. (7) were digitized, and the complex form of the measured transfer function (TFFChen) was calculated. We then derived the transfer function for the transmission line model (TFFmodel), which depends on the eight model parameters: length L and wave propagation speed c0 for each of the three line segments, and two distal model parameters (
0 and
).
Reference values for these parameters were estimated by fitting
TFFmodel to TFFChen, hereby minimizing the
(complex) sum of squared differences between TFFChen and
TFFmodel (Matlab, The Mathworks, Natic, MA). To reduce the
number of parameters in the minimization procedure, L was
set to fixed values between 10 and 30 cm for the proximal and between
15 and 45 cm for the middle and distal segments. For each set of
segment lengths, the remaining five parameters were determined using a
least square fitting algorithm. The goodness of fit, expressed as the
modulus of the root mean square (RMS) value of the difference between
TFFChen and TFFmodel, was stored for all
combinations of segment lengths. The final set of parameters, further
considered as a reference, was the one yielding the lowest |RMS|.
After assessing reference model parameters, we studied the
effect of changes in characteristic impedance (half and twice the parameter values following from fitting Chen's function) of the proximal segment and of all segments together, changes in vascular tone
, i.e, the distal wave reflection coefficient (
0 = 0
1), and the effect of a 20% variation in segment length.
Fitting Measured Radial-Carotid Pressure Transfer Functions
In vivo measurements. Applanation tonometry was performed successively at the carotid (Pcar) and radial (Prad) artery with a Millar SPT-301 penlike transducer (5) in 31 controls and 30 patients, in three different centers, and by three different operators. Patients were subjects with known coronary artery disease confirmed by wall motion abnormalities (rest and/or during dobutamine infusion) on a recent echocardiographic examination and/or a positive treadmill stress test. No patients with uncontrolled hypertension were included. The measured sequence of heartbeats was processed off-line. Using the measured electrocardiogram signal, we identified individual heart cycles, and an average pressure wave was calculated from at least five heartbeats. This signal was then calibrated by setting its mean and diastolic value equal to diastolic and mean brachial blood pressure measured with a cuff sphygmomanometer. Fourier analysis was applied on both pressure waves, and the ratio of corresponding harmonics yielded the measured transfer function (TFFmeas). A maximum of 10 harmonics were taken into account, and pressure or flow harmonics with a magnitude <1% of the first harmonic were excluded.
Fitting the model to the measured radial-carotid data.
To fit the model to the measured radial-carotid pressure transfer
function, we used Pcar as an input into the model, and the computed distal model pressure (Pd,model) was fitted to the
measured radial artery pressure (Prad) by minimization of
(Pd,model
Prad)2. We
first fitted the data by changing four parameters: the characteristic impedance of the three segments (Z0,p,
Z0,m, Z0,d) and the
modulus of the distal reflection coefficient (
0). The
remaining four parameters were given their reference value as obtained
from fitting TFFChen. Second, we further restricted the
fitting to one single parameter: the characteristic impedance of the
proximal segment (Z0,p), with the remaining
seven parameters fixed at their reference value. For both fitting
methods, we studied correlations between fitted model parameters and
(sphygmomanometer) brachial blood pressure, subject age, heart rate,
health condition (control subject or patient), and whether model
parameters were different for controls and patients.
Average TFF and Maximal and Minimal Enveloping Transfer Functions
We derived an average transfer function (TFFavg) for the whole group of 31 controls and 30 patients. Measured transfer functions were resampled at 0.25-Hz intervals. Averaging of all 61 transfer functions yielded TFFavg. Furthermore, two TFF were selected from all fitted transfer functions, representing an upper and lower enveloping curve. Because the low frequency harmonics are most important, we chose the transfer function with the highest (TFFmax) and lowest (TFFmin) amplitude at 2 Hz.TFFmax, TFFmin, and TFFavg were
used to reconstruct the carotid artery pressure (Pcar,r)
from measured Prad, with data from all three centers
included. We then calculated pulse pressure (PP = systolic
diastolic pressure) and the AIx from Pcar,r and Pcar. AIx is calculated as ±(Psys
Pinfl)/PP, with Psys being systolic pressure
and Pinfl as the pressure corresponding to the first
inflection point on the carotid pressure wave. AIx is positive when the
inflection pressure precedes systolic pressure and indicates an A-type
wave (18); it is negative when the inflection point occurs
after systolic pressure (C-type wave). We compared PP and AIx derived
from Pcar and Pcar,r with TFFmin,
TFFavg, and TFFmax using linear regression
analysis. We further calculated RMS as
|
| |
RESULTS |
|---|
|
|
|---|
Fitting Chen's Radial-Aorta Pressure Transfer Function
The agreement between the fitted model TFFmodel and TFFChen is shown in Fig. 2. Both modulus and phase angle were well predicted by the model. The derived model parameters for the different segments are given in Table 1. The characteristic impedance for proximal, middle, and distal segment was 0.44, 1.33, and 4.08 mmHg · ml
1
· s
1, respectively. The total length of the model was
66 cm. The distal reflection coefficient was practically constant
(
0 = 0.86), independent of frequency (
= 0.00).
|
With changing Z0,p (proximal segment
properties) alone, transfer functions with very distinct morphologies
were obtained (Fig. 3). A reduction of
Z0,p increases the peak of the transfer function and shifts it to lower frequencies. Because of a lower wave velocity, the time delay between the proximal and distal location increases, and
the phase angle becomes more negative. These effects are amplified with
an overall change of Z0. With
Z0 divided by 2, the TFF peak is still shifted
to lower frequencies, but there is hardly any amplification for
frequencies <3 Hz, and the higher frequencies are damped out. Vascular
tone changes are modeled as changes of the reflection coefficient.
,
being real, does not change the TFF phase angle or the location of the
peak. Only its magnitude increases with a stronger reflection
(vasoconstriction). The overall length of the path was changed by
±20%. A longer path length dampens the peak, shifts it to lower
frequencies, and increases the phase angle; the inverse is found for a
shorter path.
|
Fitting Measured Radial-Carotid Pressure Transfer Functions
Hemodynamic data for the complete population are given in Table 2 as well as average values for Z0,p, Z0,m, Z0,d, and
0, obtained from
fitting the model to measured radial-carotid transfer functions.
Overall, there was good agreement between measured and fitted radial
artery pressure (RMS = 3.3 ± 2.0 mmHg), but the fitting was
slightly better for the control than for the patient group (RMS = 2.8 ± 1.4 vs. 3.9 ± 2.5 mmHg; P < 0.05).
Z0,p and Z0,m were higher
in the patient group, whereas Z0,d was higher in
the control group, but differences were not significant.
0 was higher in the patient group (0.85 ± 0.21 vs.
0.71 ± 0.25; P < 0.05). There was no correlation
between any of the fitted model parameters and subject age,
sphygmomanometer blood pressure (systolic, diastolic, or pulse
pressure), heart rate, or the subject being classified as a patient or
control.
|
With the model fitted to the measurements using only
Z0,p, the fitting is again better in controls
(RMS = 4.6 ± 2.2 vs. 5.4 ± 3.3 mmHg).
Z0,p is higher in patients than in controls, but the difference is not statistically significant (1.10 ± 0.64 vs. 1.22 ± 0.60 mmHg · ml
1 · s
1). Again, there was no correlation between
Z0,p and subject age, sphygmomanometer blood
pressure (systolic, diastolic, or pulse pressure), heart rate, or the
subject being classified as a patient or control.
Average TFF and Maximal and Minimal Enveloping Transfer Functions
The calculated point-to-point TFFavg is shown in Fig. 4, together with TFFmax and TFFmin. The error bars indicate means ± 2 SE. TFFmax is the transfer function with the highest amplitude at 2 Hz (2.0) and is calculated using 0.22, 1.15, and 5.34 mmHg · ml
1 · s
1 for
Z0,p, Z0,m, and
Z0,d, respectively, and 0.95 for
0. All other parameters have reference values (Table 1).
TFFmin is the transfer function with the lowest amplitude
at 2 Hz (0.9) and is calculated with 1.12, 2.39, and 4.32 mmHg · ml
1 · s
1 for
Z0,p, Z0,m, and
Z0,d, respectively, and 0.51 for
0.
|
We reconstructed carotid pressure from measured radial artery pressure
using TFFmax, TFFmin, and
TFFavg, respectively. For the whole data set, the
average difference between reconstructed and measured carotid pressure,
expressed as the RMS difference, was 7.8 ± 2.7, 6.9 ± 3.4, and 4.3 ± 2.3 mmHg, respectively. The correlations between PP and
the AIx, derived from the reconstructed and measured carotid pressure,
are given in Fig. 5. Best results are
found with TFFavg, yielding an average of 41.5 ± 11.8 vs. 41.1 ± 10.0 mmHg measured. A paired t-test
indicated that both values were not significantly different
(P = 0.59). Correlations were less good for AIx.
TFFavg gave the best results, with an average value of
0.02 ± 0.19 vs. 0.01 ± 0.19 mmHg measured directly on the
carotid artery pressure wave. However, values were not significantly
different (P = 0.22; paired t-test). PP and
AIx, calculated from measured or reconstructed carotid artery pressure, were finally grouped for patients and controls (Fig.
6). PP and AIx were higher in patients
than in controls, irrespective the calculation on measured or
reconstructed pressures.
|
|
| |
DISCUSSION |
|---|
|
|
|---|
This study is based on a transmission line model for the aorta-radial pathway. The model is similar to the finger-carotid model of Karamanoglu and Fenely (9), consisting of three tube segments with the same values for tube diameters and wall viscosity. We first assessed reference model parameters by fitting the model directly to a recently invasively assessed transfer function (7, 8). The model parameters (Table 1) are within physiological ranges. The total length of the transmission line is 66 cm and is close to the average length of the arm (radial to sternum) that we measured in 13 subjects (66 ± 6 cm). Pulse-wave velocity varies between 6.5 and 8.5 m/s and is within reported ranges (9, 17). A good fit to Chen's data is obtained with a real terminal reflection coefficient, in contrast to Karamanoglu and Fenely (9). The magnitude (0.86) is higher than what Karamanoglu and Fenely (9) found at the finger level (0.64) but is close to the value of 0.8 reported for the femoral artery (15).
It was recently shown that a simple single tube with a single distal reflection site can be used to model the transfer function for "single tubelike" aortic branches such as the carotid artery and to some extent the brachial artery (21). However, such a model is too simple for the radial-aorta pathway, consisting of multiple reflection sites, and it was impossible to fit such a model to Chen's data. Using three tubes in series, three reflection sites are generated: two sites due to the impedance mismatch between the segments and a distal reflection site at the model termination. Varying the model parameters leads to logical and expected changes in the morphology of the transfer function as shown in Fig. 3 and as summarized in RESULTS.
The model was used to fit the measured radial-carotid transfer function
in 61 subjects. The fitting was first limited to
Z0,p, Z0,m,
Z0,d, and
0. Average model
parameters in the control and patient group are consistent with
reported physiopathological changes in arterial mechanical properties.
In hypertension, for instance, the stiffness of the larger elastic
arteries is increased, but the effects are far less clear for more
peripheral, muscular arteries such as the radial or femoral artery
(13, 20, 22). Nevertheless,
individually, there was no correlation between model parameters and
subject characteristics such as age, blood pressure, heart rate, or
subject classification as a control or as a patient.
A whole family of TFF curves can be obtained only by changing the characteristic impedance of the proximal segment (Fig. 3). It is further expected that changes in arterial mechanical properties are most pronounced for the larger elastic vessels close to the heart (13, 20, 22). Therefore, we studied whether the fitting could be restricted to Z0,p and whether this improved the correlation between the fitted model parameter and the patient data. Fittings were less good than with the four parameters but still acceptable (overall mean RMS of 5.0 ± 2.8 mmHg). The resulting Z0,p were comparable to the values that were found when fitting four parameters (Table 2), with the same tendency for an increased Z0,p in the patient group. However, again, there was no correlation between Z0,p and subject characteristics.
Z0,p was not different between patients and controls (Table 2). A possible explanation is the limited number of subjects included in the study. Also, blood pressure is only slightly higher in the patient group, and the difference in arterial compliance, most prominent in the ascending aorta and aortic arch, may not be high enough to yield significant differences in Z0,p.
The absence of a (simple) relation between model parameters and patient
data inhibited the computation of individualized transfer functions.
This may be due to the simplicity of the model (and the fitting
procedure), with the model parameters not reflecting physiological
properties. We also applied a model with reference parameter values
derived for the radial-aorta transfer function on measured
radial-carotid transfer functions. Still our approach is defendable.
The radial-aorta and radial-carotid pathway should at least share the
properties of the middle and distal segment; the differences exist in
the small part of the ascending aorta and the carotid artery. We
assumed the same length for these segments, but their mechanical
properties are effectively different: average Z0,p for the radial-carotid data (4 parameter
fit) is 1.08 mmHg · ml
1 · s
1
and is higher than Z0,p that was found for the
radial-aorta fit (0.44 mmHg · ml
1 · s
1). This agrees with the smaller caliber of the carotid
artery. One might think of introducing arm length as a known variable into the model. This will change the values of estimated
Z0,p, but it is unlikely that the introduction
of a parameter (radial-carotid path length), which cannot be measured
with great accuracy either, would change our observations. One should
also consider the fact that not only the total path length but also the
relative length of the three segments determines TFF.
TFFavg yielded the best approximation of PP and AIx,
whereas TFFmax or TFFmin gave (very) different
values for PP and AIx than the values derived from the directly
measured pressure wave. However, it is remarkable that when averaged
over control subjects and patients, PP and AIx were different for both
groups, irrespective the used transfer function. Thus, when
(generalized) pressure transfer functions are used in clinical studies,
comparing control subjects with patient groups or comparing different
interventions, they may potentially yield reconstructed central (aorta)
pressures and derived indexes, discriminating between groups or
interventions. Nevertheless, individually, there are considerable
differences for PP and AIx derived from the measured or reconstructed
carotid artery pressure, as indicated by the (modified) Bland-Altman
plot (3) in Fig. 7 (data
shown for TFFavg only). Furthermore, we feel that derived
parameters, such as the AIx, require reconstructed waves in which
enough high-frequency information is present to fully capture the
details of the curve. Using other algorithms, based on autoregression
models (8), may give better results than the
frequency-domain approach we used.
|
This study shows that arterial pressure waves, reconstructed using (generalized) transfer functions, carry analog information as the carotid blood pressure. This, however, does not imply that these reconstructed pressures are a better approximation of central aortic pressure than carotid blood pressure. However, in clinical settings, it is easier to monitor central pressure from an easily accessible peripheral site such as the radial artery, which is more appropriate for applanation tonometry than the carotid artery. Applanation tonometry at the carotid artery is more tedious, more time consuming, and requires skilled people and averaging of the curves. Furthermore, it is better to avoid carotid applanation tonometry in people with carotid plaque. In our study, we had to exclude about 10% of patients because no reliable carotid tonometric signals could be recorded.
We conclude that using averaged, generalized TFF leads to significant discrepancies between reconstructed and measured pressure waveforms and subsequent derived indexes. Nevertheless, very distinct transfer functions (TFFmax, TFFavg, and TFFmin) seem to preserve the information in the pressure wave that is needed for the computation of indexes as PP and the AIx, discriminating patients from healthy subjects.
| |
ACKNOWLEDGEMENTS |
|---|
This research is funded by a specialization grant of the Flemish Institute for the Promotion of the Scientific-Technological Research in Industry (IWT 943065-IWT 960250) and a concerted action program of the University of Gent, supported by the Flemish government (GOA-95003). S. Carlier was the recipient of a grant from the North Atlantic Treaty Organization and from the Belgian American Educational Foundation.
| |
FOOTNOTES |
|---|
Address for reprint requests and other correspondence: P. Segers, Hydraulics Laboratory, Institute Biomedical Technology, Univ. of Gent, Sint-Pietersnieuwstraat 41, B-9000 Gent, Belgium (E-mail: patrick.segers{at}navier.rug.ac.be).
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. §1734 solely to indicate this fact.
Received 11 March 1999; accepted in final form 4 February 2000.
| |
REFERENCES |
|---|
|
|
|---|
1.
Aakhus, S,
Soerlie C,
Faanes A,
Hauger S,
Bjoernstad K,
Hatle L,
and
Angelsen B.
Noninvasive computerized assessment of left ventricular performance and systemic hemodynamics by study of aortic root pressure and flow estimates in healthy men, and men with acute and healed myocardial infarction.
Am J Cardiol
72:
260-267,
1993[ISI][Medline].
2.
Avolio, A.
Multi-branched model of the human arterial system.
Med Biol Eng Comput
18:
709-718,
1980[ISI][Medline].
3.
Bland, JM,
and
Altman DG.
Statistical methods for assessing agreement between two methods of clinical measurement.
Lancet
1:
307-310,
1986[ISI][Medline].
4.
Cameron, JD,
McGrath B,
and
Dart A.
Use of radial artery applanation tonometry and a generalized transfer function to determine aortic pressure augmentation in subjects with treated hypertension.
J Am Coll Cardiol
32:
1214-1220,
1998
5.
Carlier, S,
Marwick T,
Segers P,
Verdonck P,
Van Camp G,
Cosyns B,
and
Thomas JD.
Non-invasive estimation of left ventricular power output by simultaneous recordings of applanation tonometry and echocardiographic Doppler signals.
In: Computers in Cardiology. IEEE Computer Society Press, 1997, p. 223-226.
6.
Chen, CH,
Ting CT,
Nussbacher A,
Nevo E,
Kass D,
Pak P,
Wang SP,
Chang MS,
and
Yin FC.
Validation of carotid artery tonometry as a means of estimating augmentation index of ascending aortic pressure.
Circulation
27:
168-175,
1996.
7.
Chen, C-H,
Nevo E,
Fetics B,
Pak PH,
Yin FCP,
Maughan L,
and
Kass DA.
Estimation of central aortic pressure waveform by mathematical transformation of radial tonometry pressure: validation of generalized transfer function.
Circulation
95:
1827-1836,
1997
8.
Fetics, B,
Nevo E,
Chen CH,
and
Kass DA.
Parametric model derivation of transfer function for noninvasive estimation of aortic pressure bt radial tonometry.
IEEE Trans Biomed Eng
46:
698-706,
1999[ISI][Medline].
9.
Karamanoglu, M,
and
Fenely M.
On-line synthesis of the human ascending aortic pressure pulse from the finger pulse.
Hypertension
30:
1416-1424,
1997
10.
Karamanoglu, M,
Gallagher D,
Avolio A,
and
O'Rourke M.
Pressure wave propagation in a multibranched model of the human upper limb.
Am J Physiol Heart Circ Physiol
269:
H1363-H1369,
1995
11.
Karamanoglu, M,
and
FM
Derivation of the ascending aorta-carotid pressure transfer function with an arterial model.
Am J Physiol Heart Circ Physiol
271:
H2399-H2404,
1996
12.
Karamanoglu, M,
O'Rourke M,
Avolio A,
and
Kelly R.
An analysis of the relationship between central aortic and peripheral upper limb pressure waves in man.
Eur Heart J
14:
160-167,
1993
13.
Kawasaki, T,
Sasayama S,
Yagi S,
Asakawa T,
and
Hirai T.
Non-invasive assessment of the age related changes in stiffness of major branches of the human arteries.
Cardiovasc Res
21:
678-687,
1987[ISI][Medline].
14.
Kelly, RP,
and
Fitchett D.
Noninvasive determination of aortic input impedance and external left ventricular power output: a validation and repeatability study of a new technique.
J Am Coll Cardiol
20:
952-963,
1992[Abstract].
15.
Li, JKJ,
Melbin J,
and
Noordergraaf A.
Directional disparity of pulse reflection in the dog.
Am J Physiol Heart Circ Physiol
247:
H95-H99,
1984.
16.
Marcus, R,
Korcarz C,
McCray G,
Neumann A,
Murphy M,
Borow K,
Weinert L,
Bednarz J,
Gretler D,
Spencer K,
Sareli P,
and
Lang R.
Noninvasive method for determination of arterial compliance using Doppler echocardiography and subclavian pulse tracings.
Circulation
89:
2688-2699,
1994
17.
Milnor, WR.
Hemodynamics. Baltimore, MD: Williams & Wilkins, 1989.
18.
Murgo, JP,
Westerhof N,
Giolma JP,
and
Altobelli SA.
Aortic input impedance in normal man: relationship to pressure wave forms.
Circulation
62:
105-116,
1980
19.
O'Rourke, M.
Arterial stiffness, systolic blood pressure, and logical treatment of arterial hypertension.
Hypertension
15:
339-347,
1990
20.
Reneman, RS,
Hoeks APG,
and
Westerhof N.
Non-invasive assessment of artery wall properties in humans-methods and interpretation.
J Vasc Invest
2:
53-64,
1996.
21.
Stergiopulos, N,
Westerhof B,
and
Westerhof N.
Physical basis of pressure transfer from periphery to aorta: a model-based study.
Am J Physiol Heart Circ Physiol
274:
H1386-H1392,
1998
22.
Van Merode, T,
Brands P,
Hoeks A,
and
Reneman R.
Different effects of ageing on elastic and muscular bifurcations in men.
J Vasc Res
33:
47-52,
1996[ISI][Medline].
23.
Westerbacka, J,
Wilkinson I,
Cockroft J,
Utriainen T,
Vehkavaara S,
and
Yji-Jarvinen H.
Diminsihed wave reflection in the aorta. A novel physiological action of insulin on large blood vessels.
Hypertension
33:
1118-1122,
1999
24.
Womersley JR. An Elastic Tube Theory of Pulse Transmission and
Oscillatory Flow in Mammalian Arteries. Wright Air Development
Centre, report no. WADC-TR56-614, 1957.
This article has been cited by other articles:
![]() |
C.-J. Thore, J. Stalhand, and M. Karlsson Toward a noninvasive subject-specific estimation of abdominal aortic pressure Am J Physiol Heart Circ Physiol, September 1, 2008; 295(3): H1156 - H1164. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Swamy, Q. Ling, T. Li, and R. Mukkamala Blind identification of the aortic pressure waveform from multiple peripheral artery pressure waveforms Am J Physiol Heart Circ Physiol, May 1, 2007; 292(5): H2257 - H2264. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. J. Stok, B. E. Westerhof, and J. M. Karemaker Changes in finger-aorta pressure transfer function during and after exercise J Appl Physiol, October 1, 2006; 101(4): 1207 - 1214. [Abstract] [Full Text] [PDF] |
||||
![]() |
P M Mottram, B A Haluska, R Leano, S Carlier, C Case, and T H Marwick Relation of arterial stiffness to diastolic dysfunction in hypertensive heart disease Heart, December 1, 2005; 91(12): 1551 - 1556. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. M. Mottram, B. Haluska, R. Leano, D. Cowley, M. Stowasser, and T. H. Marwick Effect of Aldosterone Antagonism on Myocardial Dysfunction in Hypertensive Patients With Diastolic Heart Failure Circulation, August 3, 2004; 110(5): 558 - 565. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |